The Anesthesia Classic meets with the safety Features recommended by the international standards organization.
The Anesthesia Classic has panel mounting line Pressure gauges with a drawer.
Oxygen & Nitrous Oxide flow meter tubes accurately caliberated. Oxygen flow is received downstream of other Gas, being the preferred Gas.
Incorporate one vaporizer for Halothane.
Two pairs of Yokes with Regulators & Gauges Colour coded for Oxygen and Nitrous Oxide.
To minimize backflow of Gases, non return cum pressure relief valve is provided.
Emergency Oxygen flush button cum outlet at table level.
| Structure |
Square tubular electro statically Powder coated steel sections. |
| Top Tray |
Steel top tray at eye level to keep monitoring equipments. |
| Table Top |
Stainless Steel. |
| Drawer |
Extra large one storage Drawer. |
| Cylinder Yokes |
Pin-Indexed gas specific dual Oxygen & Nitrous Oxide Yokes with gauges |
| Regulators |
For higher efficiency and reliability two each for oxygen and Nitrous Oxide. |
| Nitrous Cut-off |
Nitro cut-off system is provided which activates only when Oxygen Flowmeter control valve opens to prevent the risk of administering 100% Nitrous Oxide. |
| Oxygen failure warning device |
Pneumatic device which activates audible alarm when Oxygen supply Pressure falls 215 kpa. |
| Rotameters |
Antisatatic colour coaded non inter changeable Long (230 mm) rotating bobbin flow meters accurately calibrated.
Oxygen - 100 ml / min. to 8 Litres / min.
Nitrous Oxide - 200 ml / min to 12 Litres / min.
Oxygen flow control is on left side of flowmeter bank as it is Preferred Gas. |
| Vaporisers |
One plenum type vaporiser for Halothane supplied as standard. |
| Non-return cum Pressure relief Valve |
Minimises risk of backflow of gases, blows off when pressure exceeds 200 cm water column. |
| Emergency Oxygen |
Guarded Oxygen flush button Out-let at table level. |
| Accessories |
One Schrader type self sealing Oxygen out-lets (4.22 kg/cm2) for ventilators etc. Extended rear platform for two 10 litre water capacity cylinders. Circles absorber can be securedly Mounted on the left side of the main frame. |
| Breathing Attachment |
Standard magill's circuit. |
| Mobility |
Large diameter caster wheels (130 mm) with ball bearings. Front Casters have brakes. |
| Change over |
Anaesthesia Classic is also availble in stainless steel structure. |